System integration
Proline Promass Q 500 PROFIBUS PA
94
Hauser
Selection: control totalizer
CHANNEL
Value SETTOT
Control totalizer
33310
0
Totalize
33046
1
Resetting
33308
2
Adopt totalizer initial setting
Factory setting
Function block
Factory setting: Value SETTOT (meaning)
Totalizer 1, 2 and 3
0 (totalizing)
Data structure
Output data of SETTOT
Byte 1
Control variable 1
Input data of TOTAL
Byte 1
Byte 2
Byte 3
Byte 4
Byte 5
Measured value: floating point number (IEEE 754)
Status
SETTOT_MODETOT_TOTAL module
The module combination consists of the SETTOT, MODETOT and TOTAL functions:
• SETTOT: Control the totalizers via the PROFIBUS master.
• MODETOT: Configure the totalizers via the PROFIBUS master.
• TOTAL: Transmit totalizer value, along with the status, to the PROFIBUS master.
Three Totalizer blocks are available (slot 9 to 11).
Selection: totalizer configuration
CHANNEL
MODETOT value
Totalizer configuration
33306
0
Balancing
33028
1
Balance the positive flow
32976
2
Balance the negative flow
32928
3
Stop totalizing
Factory setting
Function block
Factory setting: Value MODETOT (meaning)
Totalizer 1, 2 and 3
0 (balancing)
Data structure
Output data of SETTOT and MODETOT
Byte 1
Byte 2
Control variable 1: SETTOT
Control variable 2: MODETOT