Technical data
Proline Promass Q 500 PROFIBUS PA
256
Hauser
o.f.s. = of full scale value
When there is a difference between the temperature at zero point adjustment and the
process temperature, the typical measured error of the sensor is
DN 25 (1"): ±0.0001 % o.f.s./°C (±0.00005 % o.f.s./°F)
DN 50, 80, 100 (2", 3", 4"): ±0.00015 % o.f.s./°C (±0.000075 % o.f.s./°F)
Density
When there is a difference between the density calibration temperature and the process
temperature, the typical measured error of the sensor is
±0.015 kg/m
3
/°C (±0.0075 kg/m
3
/°F)
[kg/m ]
3
0
0.5
1.0
1.5
2.0
2.5
3.0
[°C]
[°F]
200
150
100
50
0
-50
-100
-150
-200
-300
300
0
200
100
-100
-200
400
A0028731
Temperature
±0.005 · T °C (± 0.005 · (T – 32) °F)
Influence of medium
pressure
Mass flow
DN
[% o.r./bar]
[% o.r./psi]
[mm]
[in]
25
1
–0.004
–0.00028
50
2
–0.003
–0.00021
80
3
–0.0085
–0.00059
100
4
–0.0065
–0.00045
Volume flow
DN
[% o.r./bar]
[% o.r./psi]
[mm]
[in]
25
1
–0.0035
–0.00024
50
2
+0.0005
+0.00003
80
3
–0.008
–0.00055
100
4
–0.0065
–0.00045