Technical data
Proline Promass Q 500 PROFIBUS PA
266
Hauser
A device that has a WLAN interface (can be ordered as an option) is required for the
WLAN connection: order code for "Display; operation", option
G
"4-line, illuminated; touch
c WLAN". The device acts as an Access Point and enables communication by
computer or a mobile handheld terminal.
Supported functions
Data exchange between the operating unit (such as a notebook for example) and the
measuring device:
– Uploading the configuration from the measuring device (XML format, configuration
backup)
– Save the configuration to the measuring device (XML format, restore configuration)
– Export event list (.csv file)
– Export parameter settings (.csv file, create documentation of the measuring point
configuration)
– Export the Heartbeat verification log (PDF file, only available with the "Heartbeat
Verification" application package)
– Flash firmware version for device firmware upgrade, for instance
– Download driver for system integration
HistoROM
data management
The measuring device features HistoROM data management. HistoROM data management
comprises both the storage and import/export of key device and process data, making
operation and servicing far more reliable, secure and efficient.
When the device is delivered, the factory settings of the configuration data are stored
as a backup in the device memory. This memory can be overwritten with an updated
data record, for example after commissioning.
Additional information on the data storage concept
There are different types of data storage units in which device data are stored and used by the device:
Device memory
T-DAT
S-DAT
Available data
• Event history, such as diagnostic
events
• Parameter data record backup
• Device firmware package
• Driver for system integration e.g.:
GSD for PROFIBUS PA
• Measured value memory ("Extended
HistoROM" order option)
• Current parameter data record (used
by firmware at run time)
• Maximum indicators (min/max values)
• Totalizer values
• Sensor data: diameter etc.
• Serial number
• User-specific access code (to use the
"Maintenance" user role)
• Calibration data
• Device configuration (e.g. SW options, fixed
I/O or multi I/O)
Storage location
Fixed on the user interface board in
the connection compartment
Can be plugged into the user interface
board in the connection compartment
In the sensor plug in the transmitter neck part
Data backup
Automatic
• The most important device data (sensor and transmitter) are automatically saved in the
DAT modules
• If the transmitter or measuring device is replaced: once the T-DAT containing the
previous device data has been exchanged, the new measuring device is ready for
operation again immediately without any errors
• If the sensor is replaced: once the sensor has been replaced, new sensor data are
transferred from the S-DAT in the measuring device and the measuring device is ready
for operation again immediately without any errors
Содержание Proline Promass Q 500
Страница 279: ......
Страница 280: ...www addresses endress com...