Commissioning
Proline Promass Q 500 PROFIBUS PA
124
Hauser
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Switch-on delay
In the
Relay output function
parameter, the
Limit
option is
selected.
Define delay for the switch-on
of status output.
0.0 to 100.0 s
0.0 s
Failure mode
–
Define output behavior in
alarm condition.
• Actual status
• Open
• Closed
Open
*
Visibility depends on order options or device settings
10.6.12 Configuring the double pulse output
The
Double pulse output
submenu guides the user systematically through all the
parameters that have to be set for configuring the double pulse output.
Navigation
"Setup" menu → Double pulse output
‣
Double pulse output
Master terminal number
Slave terminal number
Signal mode
Assign pulse output 1
Measuring mode
Value per pulse
Pulse width
Failure mode
Invert output signal
Parameter overview with brief description
Parameter
Description
Selection / User interface /
User entry
Factory setting
Signal mode
Select the signal mode for the double pulse
output.
• Passive
• Active
• Passive NAMUR
Passive
Master terminal number
Shows the terminal numbers used by the
master of the double pulse output module.
• Not used
• 24-25 (I/O 2)
–
Slave terminal number
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
–