Installation
Proline Promass Q 500 PROFIBUS PA
26
Hauser
RUPTURE DISK
i
1
2
3
A0030346
1
Rupture disk label
2
Rupture disk with 1/2" NPT internal thread with 1" width across flat
3
Transport protection
Zero point adjustment
All measuring devices are calibrated in accordance with state-of-the-art technology.
Calibration takes place under reference conditions → 253. Therefore, a zero point
adjustment in the field is generally not required.
Experience shows that zero point adjustment is advisable only in special cases:
• To achieve maximum measuring accuracy even with low flow rates
• Under extreme process or operating conditions (e.g. very high process temperatures or
very high-viscosity fluids).
Protective cover
213 (8.4)
39 (1.5)
243 (9.6)
203 (8.0)
A0029552
8
Weather protection cover for Proline 500 – digital