Technical data
Proline Promass Q 500 PROFIBUS PA
248
Hauser
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
Frequency output
Maximum input values
DC 30 V, 250 mA (passive)
Maximum output current
22.5 mA (active)
Open-circuit voltage
DC 28.8 V (active)
Output frequency
Adjustable: end value frequency 2 to 10
000 Hz (f
max
= 12
500 Hz)
Damping
Adjustable: 0 to 999 s
Pulse/pause ratio
1:1
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
The range of options increases if the measuring device has one or more
application packages.
Switch output
Maximum input values
DC 30 V, 250 mA (passive)
Open-circuit voltage
DC 28.8 V (active)
Switching behavior
Binary, conductive or non-conductive
Switching delay
Adjustable: 0 to 100 s
Number of switching
cycles
Unlimited
Assignable functions
• Off
• On
• Diagnostic behavior
• Limit value
– Mass flow
– Volume flow
– Corrected volume flow
– Density
– Reference density
– Temperature
– Totalizer 1-3
• Flow direction monitoring
• Status
– Partially filled pipe detection
– Low flow cut off
The range of options increases if the measuring device has one or more
application packages.
Relay output
Function
Switch output
Version
Relay output, galvanically isolated
Switching behavior
Can be set to:
• NO (normally open), factory setting
• NC (normally closed)