Commissioning
Proline Promass Q 500 PROFIBUS PA
114
Hauser
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Operating mode
–
Define the output as a pulse,
frequency or switch output.
• Pulse
• Frequency
• Switch
Pulse
Terminal number
–
Shows the terminal numbers
used by the PFS output
module.
• Not used
• 24-25 (I/O 2)
–
Signal mode
–
Select the signal mode for the
PFS output.
• Passive
• Active
Passive
Assign pulse output 1 to n
In the
Operating mode
parameter, the
Pulse
option is
selected.
Select process variable for
pulse output.
• Off
• Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow
• Carrier mass flow
Off
Value per pulse
In the
Operating mode
parameter, the
Pulse
option is
selected and one of the
following options is selected in
the
Assign pulse output
parameter (→ 114):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
Enter measured value at which
a pulse is output.
Signed floating-point
number
Depends on country
and nominal
diameter
Pulse width
In the
Operating mode
parameter, the
Pulse
option is
selected and one of the
following options is selected in
the
Assign pulse output
parameter (→ 114):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
Define time width of the
output pulse.
0.05 to 2
000 ms
100 ms
Failure mode
In the
Operating mode
parameter, the
Pulse
option is
selected and one of the
following options is selected in
the
Assign pulse output
parameter (→ 114):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
Define output behavior in
alarm condition.
• Actual value
• No pulses
No pulses
Invert output signal
–
Invert the output signal.
• No
• Yes
No
*
Visibility depends on order options or device settings