Proline Promass Q 500 PROFIBUS PA
Commissioning
Hauser
111
Signal mode
Assign current output 1 to n
Current span
0/4 mA value
20 mA value
Fixed current
Failure mode
Failure current
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Assign current output 1 to n
–
Select process variable for
current output.
• Off
• Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow
• Carrier mass flow
• Density
• Reference density
• Concentration
• Temperature
• Carrier pipe
temperature
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
amplitude 0
• Frequency
fluctuation 0
• Oscillation
damping 0
• Oscillation
damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
Mass flow
Terminal number
–
Shows the terminal numbers
used by the current output
module.
• Not used
• 24-25 (I/O 2)
–
Current span
–
Select current range for
process value output and
upper/lower level for alarm
signal.
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
• Fixed current
Country-specific:
• 4...20 mA NAMUR
• 4...20 mA US
Signal mode
–
Select the signal mode for the
current output.
• Passive
• Active
Passive