Proline Promass Q 500 PROFIBUS PA
System integration
Hauser
93
Function block
Factory setting
AI 7
Off
AI 8
Off
Data structure
Input data of Analog Input
Byte 1
Byte 2
Byte 3
Byte 4
Byte 5
Measured value: floating point number (IEEE 754)
Status
TOTAL module
Transmit a totalizer value from the measuring device to the PROFIBUS master (Class 1).
A selected totalizer value, along with the status, is cyclically transmitted to a PROFIBUS
Master (Class 1) via the TOTAL module. The totalizer value is depicted in the first four
bytes in the form of a floating point number as per the IEEE 754 standard. The fifth byte
contains standardized status information pertaining to the totalizer value.
Three Totalizer blocks are available (slot 9 to 11).
Selection: totalizer value
The totalizer value can be specified using the CHANNEL parameter.
CHANNEL
Input variable
32961
Mass flow
33122
Volume flow
33093
Corrected volume flow
901
Target fluid mass flow
793
Carrier mass flow
1)
Only available with the "Concentration" application package
Factory setting
Function block
Factory setting: TOTAL
Totalizer 1, 2 and 3
Mass flow
Data structure
Input data of TOTAL
Byte 1
Byte 2
Byte 3
Byte 4
Byte 5
Measured value: floating point number (IEEE 754)
Status
SETTOT_TOTAL module
The module combination consists of the SETTOT and TOTAL functions:
• SETTOT: Control the totalizers via the PROFIBUS master.
• TOTAL: Transmit totalizer value, along with the status, to the PROFIBUS master.
Three Totalizer blocks are available (slot 9 to 11).