Proline Promass Q 500 PROFIBUS PA
Diagnostics and troubleshooting
Hauser
215
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
491 Current output 1 to n simulation
Deactivate simulation
–
Measured variable status
Quality
Good
Quality substatus
Ok
Coding (hex)
0x80 to 0x83
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
492 Simulation frequency output 1 to n
Deactivate simulation frequency output
–
Measured variable status
Quality
Good
Quality substatus
Ok
Coding (hex)
0x80 to 0x83
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
493 Simulation pulse output 1 to n
Deactivate simulation pulse output
–
Measured variable status
Quality
Good
Quality substatus
Ok
Coding (hex)
0x80 to 0x83
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
494 Switch output simulation 1 to n
Deactivate simulation switch output
–
Measured variable status
Quality
Good
Quality substatus
Ok
Coding (hex)
0x80 to 0x83
Status signal
C
Diagnostic behavior
Warning