Diagnostics and troubleshooting
Proline Promass Q 500 PROFIBUS PA
218
Hauser
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
830 Sensor temperature too high
Reduce ambient temp. around the sensor
housing
• Oscillation amplitude @1
• Oscillation amplitude @1
• Signal asymmetry
• Carrier mass flow
• Carrier pipe temperature
• Concentration
• Oscillation damping @1
• Oscillation damping @1
• Density
• Dynamic viscosity
• Sensor electronic
temperature (ISEM)
• Empty pipe detection
• Kinematic viscosity
• Low flow cut off
• Mass flow
• HBSI
• Pressure
• Exciter current @1
• Exciter current @1
• Oscillation frequency @1
• Oscillation frequency @1
• Reference density
• Corrected volume flow
• Oscillation damping
fluctuation @1
• Oscillation damping
fluctuation @1
• Frequency fluctuation
@1
• Frequency fluctuation
@1
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Status
• Volume flow
Measured variable status [from the factory]
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.