Diagnostics and troubleshooting
Proline Promass Q 500 PROFIBUS PA
216
Hauser
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
495 Diagnostic event simulation
Deactivate simulation
–
Measured variable status
Quality
Good
Quality substatus
Ok
Coding (hex)
0x80 to 0x83
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
496 Status input simulation
Deactivate simulation status input
–
Measured variable status
Quality
Good
Quality substatus
Ok
Coding (hex)
0x80 to 0x83
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
497 Simulation block output
Deactivate simulation
–
Measured variable status
Quality
Good
Quality substatus
Ok
Coding (hex)
0x80 to 0x83
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
520 I/O 1 to n hardware configuration invalid
1. Check I/O hardware configuration
2. Replace wrong I/O module
3. Plug the module of double pulse output
on correct slot
–
Measured variable status
Quality
Good
Quality substatus
Ok
Coding (hex)
0x80 to 0x83
Status signal
F
Diagnostic behavior
Alarm