Proline Promass Q 500 PROFIBUS PA
Commissioning
Hauser
117
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Maximum frequency value
In the
Operating mode
parameter, the
Frequency
option is selected and one of
the following options is
selected in the
Assign
frequency output
parameter
(→ 116):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
• Density
• Reference density
• Concentration
• Temperature
• Carrier pipe temperature
• Electronic temperature
• Oscillation frequency 0
• Frequency fluctuation 0
• Oscillation amplitude 0
• Oscillation damping 0
• Oscillation damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
Enter maximum frequency.
0.0 to 10
000.0 Hz
10
000.0 Hz
Measuring value at minimum
frequency
In the
Operating mode
parameter, the
Frequency
option is selected and one of
the following options is
selected in the
Assign
frequency output
parameter
(→ 116):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
• Density
• Reference density
• Concentration
• Temperature
• Carrier pipe temperature
• Electronic temperature
• Oscillation frequency 0
• Frequency fluctuation 0
• Oscillation amplitude 0
• Oscillation damping 0
• Oscillation damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
Enter measured value for
minmum frequency.
Signed floating-point
number
Depends on country
and nominal
diameter