Proline Promass Q 500 PROFIBUS PA
Technical data
Hauser
245
16.3 Input
Measured variable
Direct measured variables
• Mass flow
• Density
• Temperature
Calculated measured variables
• Volume flow
• Corrected volume flow
• Reference density
Measuring range
Measuring ranges for liquids
DN
Measuring device
DN
Compatible pipe diameter
Measuring range full scale
values
min(F)
to
max(F)
[mm]
[in]
[mm]
[in]
[kg/h]
[lb/min]
25
1
25/40
1/1½
0 to 20
000
0 to 736
50
2
50/80
2/3
0 to 80
000
0 to 2
944
80
3
80/100
3/4
0 to 200
000
0 to 7
360
100
4
100/150
4/6
0 to 550
000
0 to 20
240
Measuring ranges for gases
The full scale values depend on the density of the gas and can be calculated with the
formula below:
max(G)
=
max(F)
· ρ
G
: x
max(G)
Maximum full scale value for gas [kg/h]
max(F)
Maximum full scale value for liquid [kg/h]
max(G)
<
max(F)
max(G)
can never be greater than
max(F)
ρ
G
Gas density in [kg/m³] at operating conditions
x
Constant dependent on nominal diameter
DN
x
[mm]
[in]
[kg/m
3
]
25
1
100
50
2
100
80
3
120
100
4
200
Calculation example for gas
• Sensor: Promass Q, DN 50
• Gas: Air with a density of 60.3 kg/m³ (at 20 °C and 50 bar)
• Measuring range (liquid): 80
000 kg/h
• x = 100 kg/m³ (for Promass Q, DN 50)
Maximum possible full scale value:
max(G)
=
max(F)
· ρ
G
: x = 80
000 kg/h · 60.3 kg/m³ : 100 kg/m³ = 48
240 kg/h