Commissioning
Proline Promass Q 500 PROFIBUS PA
144
Hauser
Parameter
Prerequisite
Description
Selection / User
entry / User
interface
Factory setting
Pulse output simulation 1 to n
In the
Operating mode
parameter, the
Pulse
option is
selected.
Set and switch off the pulse
output simulation.
For
Fixed value
option:
Pulse width
parameter
(→ 114) defines the
pulse width of the pulses
output.
• Off
• Fixed value
• Down-counting
value
Off
Pulse value 1 to n
In the
Pulse output
simulation 1 to n
parameter,
the
Down-counting value
option is selected.
Enter the number of pulses for
simulation.
0 to 65
535
0
Switch output simulation 1 to n
In the
Operating mode
parameter, the
Switch
option
is selected.
Switch the simulation of the
switch output on and off.
• Off
• On
Off
Switch status 1 to n
–
Select the status of the status
output for the simulation.
• Open
• Closed
Open
Relay output 1 to n simulation
–
Switch simulation of the relay
output on and off.
• Off
• On
Off
Switch status 1 to n
The
On
option is selected in
the
Switch output simulation
1 to n
parameter parameter.
Select status of the relay
output for the simulation.
• Open
• Closed
Open
Pulse output simulation
–
Set and switch off the pulse
output simulation.
For
Fixed value
option:
Pulse width
parameter
defines the pulse width
of the pulses output.
• Off
• Fixed value
• Down-counting
value
Off
Pulse value
In the
Pulse output
simulation
parameter, the
Down-counting value
option
is selected.
Set and switch off the pulse
output simulation.
0 to 65
535
0
Device alarm simulation
–
Switch the device alarm on and
off.
• Off
• On
Off
Diagnostic event category
–
Select a diagnostic event
category.
• Sensor
• Electronics
• Configuration
• Process
Process
Diagnostic event simulation
–
Select a diagnostic event to
simulate this event.
• Off
• Diagnostic event
picklist (depends
on the category
selected)
Off
Logging interval
–
Define the logging interval tlog
for data logging. This value
defines the time interval
between the individual data
points in the memory.
1.0 to 3
600.0 s
–
*
Visibility depends on order options or device settings