If operating outdoors:
Avoid direct sunlight, particularly in warm climatic regions.
You can order a weather protection cover from Hauser : → 241
Vibrations
The operational reliability of the measuring system is not affected by plant vibrations.
6.1.3
Special mounting instructions
Drainability
The measuring tubes can be completely drained and protected against solids build-up in
vertical orientation.
Rupture disk
Information that is relevant to the process: → 260.
L
WARNING
Limited functional reliability of the rupture disk.
Danger to persons from escaping fluids!
‣
Do not remove the rupture disk.
‣
When using a rupture disk, do not use a heating jacket.
‣
Make sure that the function and operation of the rupture disk is not impeded through
the installation of the device.
‣
Take precautions to prevent damage and danger to persons if the rupture disk is
actuated.
‣
Observe information on the rupture disk sticker.
The position of the rupture disk is indicated on a sticker beside it.
The transportation guard must be removed.
For information on the dimensions: see the "Mechanical construction" section
The existing connecting nozzles are not intended for the purpose of rinsing or pressure
monitoring, but instead serve as the mounting location for the rupture disk.
In the event of a failure of the rupture disk, a discharge device can be screwed onto the
internal thread of the rupture disk in order to drain off any escaping medium.