Proline Promass Q 500 PROFIBUS PA
Diagnostics and troubleshooting
Hauser
173
Diagnostic information pertaining to the electronics: diagnostic number 200 to 399
Diagnostic behavior
(configurable)
Measured value status (fixed assignment)
Device diagnosis
(fixed assignment)
Quality
Quality
Substatus
Coding
(hex)
Category
(NE107)
Alarm
BAD
Maintenance
alarm
0x24 to 0x27
F
(Failure)
Maintenance
alarm
Warning
Logbook entry only
GOOD
ok
0x80 to 0x8E
–
–
Off
Diagnostic information pertaining to the configuration: diagnostic number 400 to 599
Diagnostic behavior
(configurable)
Measured value status (fixed assignment)
Device diagnosis
(fixed assignment)
Quality
Quality
Substatus
Coding
(hex)
Category
(NE107)
Alarm
BAD
Process
related
0x28 to 0x2B
F
(Failure)
Invalid process
condition
Warning
UNCERTA
IN
Process
related
0x78 to 0x7B
S
(Out of
specification)
Invalid process
condition
Logbook entry only
GOOD
ok
0x80 to 0x8E
–
–
Off
Diagnostic information pertaining to the process: diagnostic number 800 to 999
Diagnostic behavior
(configurable)
Measured value status (fixed assignment)
Device diagnosis
(fixed assignment)
Quality
Quality
Substatus
Coding
(hex)
Category
(NE107)
Alarm
BAD
Process
related
0x28 to 0x2B
F
(Failure)
Invalid process
condition
Warning
UNCERTA
IN
Process
related
0x78 to 0x7B
S
(Out of
specification)
Invalid process
condition
Logbook entry only
GOOD
ok
0x80 to 0x8E
–
–
Off
12.7 Overview of diagnostic information
In the case of some items of diagnostic information, the diagnostic behavior can be