System integration
Proline Promass Q 500 PROFIBUS PA
92
Hauser
AI module (Analog Input)
Transmit an input variable from the measuring device to the PROFIBUS master (Class 1).
The selected input variable, along with the status, is cyclically transmitted to the
PROFIBUS Master (Class 1) via the AI module. The input variable is depicted in the first
four bytes in the form of a floating point number as per the IEEE 754 standard. The fifth
byte contains standardized status information pertaining to the input variable.
Eight Analog Input blocks are available (slot 1 to 8).
Selection: input variable
The input variable can be specified using the CHANNEL parameter.
CHANNEL
Input variable
32961
Mass flow
33122
Volume flow
33093
Corrected volume flow
901
Target fluid mass flow
793
Carrier mass flow
32850
Density
33092
Reference density
794
Concentration
33101
Temperature
263
Carrier tube temperature
1042
Electronic temperature
1066
Oscillation frequency 0
1124
Oscillation amplitude 0
1062
Frequency fluctuation 0
1117
Oscillation damping 0
1054
Tube damping fluctuation 0
1056
Excitation current 0
1125
Signal asymmetry
2285
Current output 1
2286
Current output 2
2287
Current output 3
1)
Only available with the Concentration application package
2)
Only available with the Heartbeat Verification application package
Factory setting
Function block
Factory setting
AI 1
Mass flow
AI 2
Volume flow
AI 3
Corrected volume flow
AI 4
Density
AI 5
Reference density
AI 6
Temperature