Technical data
Proline Promass Q 500 PROFIBUS PA
254
Hauser
Base accuracy
Mass flow and volume flow (liquids)
±0.05 % o.r. (PremiumCal; order code for "Calibration flow", option
D
, for mass flow)
±0.10 % o.r.
Mass flow (gases)
±0.35 % o.r.
Density (liquids)
±0.2 kg/m³ / ±0.0002 g/cm³
Valid between 20 °C and 60 °C. The measured error increases by 0.015 kg/(m
3
⋅°C) outside
the temperature range.
Valid range for density calibration: 0 to 2
000 kg/m³, +20 to +60 °C (+68 to +140 °F)
The density specification only applies as of a minimum flow velocity of 0.2 m/s in relation
to the nominal diameter.
For the highest density measurement accuracy avoid significant tensile stresses due to the
installation
Temperature
±0.1 °C ± 0.003 · T °C (±0.18 °F ± 0.003 · (T – 32) °F)
Zero point stability
DN
Zero point stability
[mm]
[in]
[kg/h]
[lb/min]
25
1
0.36
0.013
50
2
1.8
0.066
80
3
5.4
0.20
100
4
11.5
0.42
Flow values
Flow values as turndown parameter depending on nominal diameter.
SI units
DN
1:1
1:10
1:20
1:50
1:100
1:500
[mm]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
25
20
000
2
000
1
000
400
200
40
50
80
000
8
000
4
000
1
600
800
160
80
200
000
20
000
10
000
4
000
2
000
400
100
550
000
55
000
27
500
11
000
5
500
1
100
Содержание Proline Promass Q 500