Assign pulse output 1
Select process variable for pulse output.
• Off
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
Off
Measuring mode
Select measuring mode for pulse output.
• Forward flow
• Forward/Reverse flow
• Reverse flow
• Reverse flow compensation
Forward flow
Value per pulse
Enter measured value at which a pulse is
output.
Signed floating-point number
Depends on country and
nominal diameter
Pulse width
Define time width of the output pulse.
0.5 to 2
000 ms
0.5 ms
Failure mode
Define output behavior in alarm condition.
• Actual value
• No pulses
No pulses
Invert output signal
Invert the output signal.
• No
• Yes
No
*
Visibility depends on order options or device settings
10.6.13 Configuring the local display
The
Display
wizard guides you systematically through all the parameters that can
configured for configuring the local display.
Navigation
"Setup" menu → Display
‣
Display
Format display
Value 1 display
0% bargraph value 1
100% bargraph value 1
Value 2 display
Value 3 display
0% bargraph value 3
100% bargraph value 3
Value 4 display