Proline t-mass I 300 HART
Diagnostics and troubleshooting
Hauser
147
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of configuration
330
Flash file invalid
1. Update firmware of device
2. Restart device
M
Warning
331
Firmware update failed
1. Update firmware of device
2. Restart device
F
Warning
410
Data transfer
1. Check connection
2. Retry data transfer
F
Alarm
412
Processing download
Download active, please wait
C
Warning
431
Trim 1 to n
Carry out trim
C
Warning
437
Configuration
incompatible
Restart device
F
Alarm
438
Dataset
1. Check data set file
2. Check device configuration
3. Up- and download new
configuration
M
Warning
441
Current output 1 to n
1. Check process
2. Check current output settings
S
Warning
442
Frequency output 1 to n
1. Check process
2. Check frequency output settings
S
Warning
443
Pulse output 1 to n
1. Check process
2. Check pulse output settings
S
Warning
444
Current input 1 to n
1. Check process
2. Check current input settings
S
Warning
453
Flow override
Deactivate flow override
C
Warning
484
Failure mode simulation
Deactivate simulation
C
Alarm
485
Measured variable
simulation
Deactivate simulation
C
Warning
486
Current input 1 to n
simulation
Deactivate simulation
C
Warning
491
Current output 1 to n
simulation
Deactivate simulation
C
Warning
492
Simulation frequency
output 1 to n
Deactivate simulation frequency
output
C
Warning
493
Simulation pulse output
1 to n
Deactivate simulation pulse output
C
Warning
494
Switch output simulation
1 to n
Deactivate simulation switch output C
Warning
495
Diagnostic event
simulation
Deactivate simulation
C
Warning
496
Status input simulation
Deactivate simulation status input
C
Warning
520
I/O 1 to n hardware
configuration invalid
1. Check I/O hardware
configuration
2. Replace wrong I/O module
3. Plug the module of double pulse
output on correct slot
F
Alarm
537
Configuration
1. Check IP addresses in network
2. Change IP address
F
Warning
539
Flow computer
configuration incorrect
1. Check input value (pressure,
temperature)
2. Check allowed values of the
medium properties
S
Alarm
Содержание HART Proline t-mass I 300
Страница 201: ......
Страница 202: ...www addresses endress com 71528699 71528699 ...