Proline t-mass I 300 HART
Commissioning
Hauser
87
Parameter
Description
Selection
Factory setting
Temperature unit
Select temperature unit.
Unit choose list
Country-specific:
• °C
• °F
Pressure unit
Select process pressure unit.
Unit choose list
Country-specific:
• bar a
• psi a
Velocity unit
Select velocity unit.
Unit choose list
Country-specific:
• m/s
• ft/s
Length unit
Select length unit for nominal diameter.
Unit choose list
Country-specific:
• mm
• in
Date/time format
Select date and time format.
• dd.mm.yy hh:mm
• dd.mm.yy hh:mm am/pm
• mm/dd/yy hh:mm
• mm/dd/yy hh:mm am/pm
–
9.4.7
Displaying the I/O configuration
The
I/O configuration
submenu guides the user systematically through all the parameters
in which the configuration of the I/O modules is displayed.
Navigation
"Setup" menu → I/O configuration
‣
I/O configuration
I/O module 1 to n terminal numbers
I/O module 1 to n information
I/O module 1 to n type
Apply I/O configuration
I/O alteration code
Parameter overview with brief description
Parameter
Description
User interface / Selection / User entry
I/O module 1 to n terminal numbers Shows the terminal numbers used by the I/O module.
• Not used
• 26-27 (I/O 1)
• 24-25 (I/O 2)
• 22-23 (I/O 3)
I/O module 1 to n information
Shows information of the plugged I/O module.
• Not plugged
• Invalid
• Not configurable
• Configurable
• HART