Diagnostics and troubleshooting
Proline t-mass I 300 HART
138
Hauser
LED
Color
Meaning
2
Device status (normal
operation)
Off
Firmware error
Green
Device status is ok.
Flashing green
Device is not configured.
Red
A diagnostic event with "Alarm" diagnostic behavior
has occurred.
Flashing red
A diagnostic event with "Warning" diagnostic
behavior has occurred.
Flashing red/green
The device restarts.
2
Device status (during
start-up)
Flashes red slowly
If > 30 seconds: problem with the boot loader.
Flashes red quickly
If > 30 seconds: compatibility problem when
reading the firmware.
3
Not used
–
–
4
Communication
Off
Communication not active.
White
Communication active.
5
Service interface (CDI)
Off
Not connected or no connection established.
Yellow
Connected and connection established.
Flashing yellow
Service interface active.