Technical data
Proline t-mass I 300 HART
170
Hauser
Damping
Configurable: 0 to 999.9 s
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• FAD volume flow
• Flow velocity
• Temperature
• Energy flow
• Pressure
• Density
• Heat flow
• Electronics temperature
• 2nd temperature heat difference
For SIL (application package), only mass flow
Current output 4 to 20 mA
Order code
"Output; input 2" (21), "Output; input 3" (022):
Option B: current output 4 to 20 mA
Signal mode
Can be set to:
• Active
• Passive
Current span
Can be set to:
• 4 to 20 mA NAMUR
• 4 to 20 mA US
• 4 to 20 mA
• 0 to 20 mA (only if the signal mode is active)
• Fixed current
Maximum output values
22.5 mA
Open-circuit voltage
DC 28.8 V (active)
Maximum input voltage
DC 30 V (passive)
Load
0 to 700 Ω
Resolution
0.38 µA
Damping
Configurable: 0 to 999.9 s
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• FAD volume flow
• Flow velocity
• Temperature
• Energy flow
• Pressure
• Density
• Heat flow
• Electronics temperature
• 2nd temperature heat difference
For SIL (application package), only mass flow
Pulse/frequency/switch output
Function
Can be set to pulse, frequency or switch output
Version
Open collector
Can be set to:
• Active
• Passive
• Passive NAMUR
Maximum input values
DC 30 V, 250 mA (passive)