Commissioning
Proline t-mass I 300 HART
124
Hauser
2. Enter the reset code.
The access code has been reset to the factory setting
0000
. It can be redefined
9.8.2
Write protection via write protection switch
Unlike parameter write protection via a user-specific access code, this allows write access
to the entire operating menu - except for the
"Contrast display" parameter
- to be locked.
The parameter values are now read only and cannot be edited any more (exception
"Contrast display" parameter
):
• Via local display
• Via HART protocol
1.
ON
OFF
1
2
3
4
A0029630
Setting the write protection (WP) switch on the main electronics module to the
ON
position enables hardware write protection.
In the
Locking status
parameter the
Hardware locked
option is displayed
→ 125. In addition, on the local display the -symbol appears in front of the
parameters in the header of the operational display and in the navigation view.
X X X X
X X X
X
X
20.50
XX
XX
A0029425
2. Setting the write protection (WP) switch on the main electronics module to the
OFF
position (factory setting) disables hardware write protection.
No option is displayed in the
Locking status
parameter → 125. On the local
display, the -symbol disappears from in front of the parameters in the header
of the operational display and in the navigation view.