Proline t-mass I 300 HART
Commissioning
Hauser
83
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User entry
Reference conditions
–
Select reference conditions for
calculation of the corrected volume
flow.
• 1013.25 mbara, 0 °C
• 1013.25 mbara, 15 °C
• 1013.25 mbara, 20 °C
• 1013.25 mbara, 25 °C
• 1000 mbara, 0 °C
• 1000 mbara, 15 °C
• 1000 mbara, 20 °C
• 1000 mbara, 25 °C
• 14.696 psia, 59 °F
• 14.696 psia, 60 °F
• User-defined
Reference pressure
The
Others
option is selected in the
Reference conditions
parameter.
Select reference conditions for the
corrected volume flow.
0 to 250 bar a
Reference temperature
–
Enter reference temperature for
calculating the reference density.
–200 to 450 °C
FAD conditions
The
Air or compressed air
option is
selected in the
Measurement
application
parameter parameter.
Select reference conditions for the
calculation of the FAD density (FAD =
free air delivery).
• 1000 mbara, 20 °C
• 14.504 psia, 68 °F
• User-defined
FAD pressure
• The
Air or compressed air
option is
selected in the
Measurement
application
parameter parameter.
• The
User-defined
option is selected
in the
FAD conditions
parameter
parameter.
Enter reference pressure for the
calulation of the FAD density (FAD =
free air delivery).
0 to 250 bar a
FAD temperature
• The
Air or compressed air
option is
selected in the
Measurement
application
parameter parameter.
• The
User-defined
option is selected
in the
FAD conditions
parameter
parameter.
Enter reference temperature for the
calulation of the FAD density (FAD =
free air delivery).
–200 to 450 °C
Reference combustion temperature
The
Energy
option is selected in the
Measurement application
parameter
parameter.
Enter reference combustion
temperature to calculate the natural
gas energy value.
–200 to 450 °C
9.4.4
Sensor adjustment
Parameters pertaining to the pipe shape of the insertion version can be configured in the
Sensor adjustment
submenu.
The number of submenus and parameters can vary depending on the device version.
Certain submenus and parameters in these submenus are not described in the
Operation Instructions. Instead a description is provided in the Special Documentation
for the device (→ "Supplementary documentation" section).
‣
Sensor adjustment
Installation direction
Installation factor
Pipe shape
Pipe inner diameter