Installation
Proline t-mass I 300 HART
22
Hauser
20 × DN
2 × DN
A0040193
4
Reduction
30 × DN
2 × DN
A0040192
5
Expansion
20 × DN
2 × DN
A0039440
6
90° elbow
25 × DN
2 × DN
A0039441
7
2 × 90° elbow
50 × DN
2 × DN
A0039445
8
Control valve
35 × DN
2 × DN
A0039442
9
2 × 90 ° elbow, 3-dimensional
%
DN
40
35
30
25
20
15
10
5
0
1
2
3
4
5
6
7
A
B
1
2
3
4
A0045846
10
The additional measured error to be expected without flow conditioners depending on the disturbance
type and inlet run
A
Additional measured error (%)
B
Inlet run (DN)
1
2 × 90 ° elbow, 3-dimensional
2
Expansion
3
2 × 90° elbow
4
Reduction or 90° elbow
Flow conditioner
Use flow conditioners if it is not possible to observe the required inlet runs. Flow
conditioners improve the flow profile and therefore reduce the necessary inlet runs.