Technical data
Proline t-mass I 300 HART
180
Hauser
15.9 Process
Medium temperature range
Sensor
–40 to +180 °C (–40 to +356 °F)
Seals
• Sealing rings:
• EPDM –40 to +140 °C (–40 to +284 °F)
• FKM –40 to +180 °C (–40 to +356 °F)
• Ferrule:
• PEEK –40 to +140 °C (–40 to +284 °F)
• PVDF –20 to +110 °C (–4 to +230 °F)
• 1.4404 –40 to +180 °C (–40 to +356 °F)
Ferrule in 1.4404: ferrule has a fixed position on the shaft. Restriction for repeat
calibrations (observe minimum insertion depth → 20)
Medium pressure range
Minimum 0.5 bar absolute. Maximum permitted medium pressure → 180
Pressure/temperature
ratings
The following pressure/temperature diagrams apply to all pressure-bearing parts of the
device and not just the process connection. The diagrams show the maximum permissible
medium pressure depending on the specific medium temperature.
Ferrule
PEEK
1.4404
-80
0
80
160
240
320
400
0
10
20
30
40
0
200
400
600
[°C]
[°F]
[bar]
[psi]
200
150
100
50
0
-50
50
A0041035-EN
44
With compression fitting material 1.4404/F316L/F316
PVDF
-80
0
80
160
240
320
400
0
10
20
30
0
200
400
[°C]
[°F]
[bar]
[psi]
200
150
100
50
0
-50
A0041590-EN
45
With compression fitting material 1.4404/F316L/F316
Flow limit
Measuring range → 164
Содержание HART Proline t-mass I 300