Proline t-mass I 300 HART
Technical data
Hauser
171
Open-circuit voltage
DC 28.8 V (active)
Voltage drop
For 22.5 mA: ≤ DC 2 V
Pulse output
Maximum input values
DC 30 V, 250 mA (passive)
Maximum output current
22.5 mA (active)
Open-circuit voltage
DC 28.8 V (active)
Pulse width
Configurable: 0.05 to 2
000 ms
Maximum pulse rate
10
000 Impulse/s
Pulse value
Adjustable
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• FAD volume flow
• Energy flow
• Heat flow
For SIL (application package), only mass flow
Frequency output
Maximum input values
DC 30 V, 250 mA (passive)
Maximum output current
22.5 mA (active)
Open-circuit voltage
DC 28.8 V (active)
Output frequency
Adjustable: end value frequency 2 to 10
000 Hz (f
max
= 12
500 Hz)
Damping
Configurable: 0 to 999.9 s
Pulse/pause ratio
1:1
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• FAD volume flow
• Flow velocity
• Temperature
• Energy flow
• Pressure
• Density
• Heat flow
• Electronics temperature
• 2nd temperature heat difference
For SIL (application package), only mass flow
Switch output
Maximum input values
DC 30 V, 250 mA (passive)
Open-circuit voltage
DC 28.8 V (active)
Switching behavior
Binary, conductive or non-conductive
Switching delay
Configurable: 0 to 100 s