Diagnostics and troubleshooting
Proline t-mass I 300 HART
146
Hauser
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
262
Sensor electronic
connection faulty
1. Check or replace connection cable
between sensor electronic module
(ISEM) and main electronics
2. Check or replace ISEM or main
electronics
F
Alarm
270
Main electronic failure
Change main electronic module
F
Alarm
271
Main electronic failure
1. Restart device
2. Change main electronic module
F
Alarm
272
Main electronic failure
Restart device
F
Alarm
273
Main electronic failure
Change electronic
F
Alarm
275
I/O module 1 to n
defective
Change I/O module
F
Alarm
276
I/O module 1 to n faulty
1. Restart device
2. Change I/O module
F
Alarm
281
Electronic initialization
Firmware update active, please wait! F
Alarm
283
Memory content
Reset device
F
Alarm
283
Memory content
Restart device
F
Alarm
302
Device verification in
progress
Device verification active, please
wait.
C
Warning
303
I/O 1 to n configuration
changed
1. Apply I/O module configuration
(parameter 'Apply I/O
configuration')
2. Afterwards reload device
description and check wiring
M
Warning
311
Electronic failure
1. Do not reset device
2. Contact service
M
Warning
332
Writing in HistoROM
backup failed
Replace user interface board
Ex d/XP: replace transmitter
F
Alarm
361
I/O module 1 to n faulty
1. Restart device
2. Check electronic modules
3. Change I/O Modul or main
electronics
F
Alarm
372
Sensor electronic (ISEM)
faulty
1. Restart device
2. Check if failure recurs
3. Replace sensor electronic module
(ISEM)
F
Alarm
373
Sensor electronic (ISEM)
faulty
Transfer data or reset device
F
Alarm
375
I/O- 1 to n
communication failed
1. Restart device
2. Check if failure recurs
3. Replace module rack inclusive
electronic modules
F
Alarm
378
Supply voltage ISEM
faulty
Check supply voltage to the ISEM
F
Alarm
382
Data storage
1. Insert T-DAT
2. Replace T-DAT
F
Alarm
383
Memory content
1. Restart device
2. Delete T-DAT via 'Reset device'
parameter
3. Replace T-DAT
F
Alarm
387
HistoROM data faulty
Contact service organization
F
Alarm