Diagnostics and troubleshooting
Proline t-mass I 300 HART
148
Hauser
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
594
Relay output simulation
Deactivate simulation switch output C
Warning
Diagnostic of process
803
Current loop
1. Check wiring
2. Change I/O module
F
Alarm
832
Electronic temperature
too high
Reduce ambient temperature
S
Warning
833
Electronic temperature
too low
Increase ambient temperature
S
Warning
834
Process temperature too
high
Reduce process temperature
S
Warning
835
Process temperature too
low
Increase process temperature
S
Warning
842
Process limit
Low flow cut off active!
1. Check low flow cut off
configuration
S
Warning
882
Input signal
1. Check input configuration
2. Check external device or process
conditions
F
Alarm
941
Flow velocity too high
1. Check process cond.
2. Increase system pressure
S
Alarm
961
Delta temperature
Check flow rate
S
Alarm
976
Mass flow out of
calibrated range
1. Check process cond.
2. Increase system pressure
S
Warning
977
Reverse flow detected
Check flow direction
S
Warning
979
Unstable process
conditions
1. Check process cond.
2. Increase system pressure
S
Warning
1)
Diagnostic behavior can be changed.
11.8 Pending diagnostic events
The
Diagnostics
menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
• Via "FieldCare" operating tool → 144
• Via "DeviceCare" operating tool → 144
Other pending diagnostic events can be displayed in the
Diagnostic list
submenu
Navigation
"Diagnostics" menu
Diagnostics
Actual diagnostics