Installation
Proline Promass 83
24
Hauser
3.3.3
Turning the local display
1.
Unscrew cover of the electronics compartment from the transmitter housing.
2.
Press the side latches on the display module and remove the module from the
electronics compartment cover plate.
3.
Rotate the display to the desired position (max. 4 × 45 ° in both directions), and reset it
onto the electronics compartment cover plate.
4.
Screw the cover of the electronics compartment firmly back onto the transmitter
housing.
a0003236
Fig. 20:
Turning the local display (field housing)
3.4
Post-installation check
Perform the following checks after installing the measuring device in the pipe:
4 x 45°
Device condition and specifications
Notes
Is the device damaged (visual inspection)?
–
Does the device correspond to specifications at the measuring point, including process
temperature and pressure, ambient temperature, measuring range, etc.?
Installation instructions
Notes
Does the arrow on the sensor nameplate match the direction of flow through the pipe?
–
Are the measuring point number and labeling correct (visual inspection)?
–
Is the orientation chosen for the sensor correct, in other words suitable for sensor type, fluid
properties (outgassing, with entrained solids) and fluid temperature?
Process environment / process conditions
Notes
Is the measuring device protected against moisture and direct sunlight?
–