Commissioning
Proline Promass 83
82
Hauser
6484
MEASURE FLUID 1
Use
P
to select START and press
F
.
The message "DENSITY MEASUREMENT RUNNING" appears on the
display for approximately 10 seconds.
During this time Promass measures the current density of the first
fluid (measured density value).
▾
For 2-point density adjustment only:
6485
DENSITY SET VALUE 2
Use
P
to enter the target density of the second fluid and press
F
to
save this value (input range = actual density value ±10%).
6486
MEASURE FLUID 2
Use
P
to select START and press
F
.
The message "DENSITY MEASUREMENT RUNNING" appears on the
display for approximately 10 seconds.
During this time Promass measures the current density of the
second fluid (measured density value).
▾
6487
DENSITY ADJUSTMENT
Use
P
to select DENSITY ADJUSTMENT and press
F
.
Promass compares the measured density value and the target
density value and calculates the new density coefficient.
6488
RESTORE ORIGINAL
If the density adjustment does not complete correctly, you can
select the RESTORE ORIGINAL function to reactivate the default
density coefficient.
▾
Back to the HOME position:
→
Press and hold down Esc key (
X
) for longer than three seconds or
→
Repeatedly press and release Esc key (
X
)
→
Exit the function matrix step by step
Function
No.
Function name
Setting to be selected (
O
or
S
)
(to next function with
F
)
Содержание HART Proline Promass 83