Proline Promass 83
Technical data
Hauser
117
Influence of medium temperature
When there is a difference between the temperature for zero point adjustment and the
process temperature, the typical measured error of the sensor is ±0.0002% of the full scale
value / °C (±0.0001% of the full scale value/°F).
Influence of medium pressure
The table below shows the effect on accuracy of mass flow due to a difference between
calibration pressure and process pressure.
Design fundamentals
Dependent on the flow:
• Flow
≥
Zero point stability ÷ (Base accuracy ÷ 100)
– Max. measured error: ±Base accuracy in % o.r.
– Repeatability: ± ½ · Base accuracy in % o.r.
• Flow < Zero point stability ÷ (Base accuracy ÷ 100)
– Max. measured error: ± (Zero point stability ÷ measured value) · 100% o.r.
– Repeatability: ± ½ · (Zero point stability ÷ measured value) · 100% o.r.
DN
Promass F (standard)
Promass F (high-temperature version)
[mm]
[inch]
[% o.r./bar]
[% o.r./bar]
8
3/8
no influence
–
15
½
no influence
–
25
1
no influence
no influence
40
1½
–0.003
–
50
2
–0.008
–0.008
80
3
–0.009
–0.009
100
4
–0.007
–
150
6
–0.009
–
250
10
–0.009
–
Base accuracy for
Mass flow liquids, PremiumCal
0.05
Mass flow liquids
0.10
Volume flow liquids
0.10
Mass flow gases
0.35
Содержание HART Proline Promass 83
Страница 2: ...Proline Promass 83 2 Endress Hauser...
Страница 147: ...Proline Promass 83 Endress Hauser 147...
Страница 148: ...www addresses endress com...