Operation
Proline Promass 83
52
Hauser
HART 7
Byte-bit
Error No.
Short error description
0-0
001
Serious device error
0-1
011
Measuring amplifier has faulty EEPROM
0-2
012
Error when accessing data of the measuring amplifier EEPROM
0-3
031
S-DAT: Defective
0-4
032
S-DAT: missing
0-5
041
T-DAT: Defective or missing
0-6
042
T-DAT: Error accessing saved values
0-7
111
Totalizer checksum error
1-0
205
T-DAT: Data download not successful
1-1
206
T-DAT: Data upload not successful
1-2
251
Internal communication fault on the amplifier board.
1-3
261
No data reception between amplifier and I/O board
1-5
355
Frequency output:
The actual value for the flow lies outside the set limits.
1-6
356
1-7
357
2-0
358
2-1
359
Pulse output:
Pulse output frequency is out of range.
2-2
360
2-3
361
2-4
362
2-5
379
The measuring tube oscillation frequency is outside the permitted range.
2-6
380
2-7
381
The temperature sensor on the measuring tube is likely defective.
3-0
382
3-1
383
The temperature sensor on the carrier tube is likely defective.
3-2
384
3-3
385
One of the measuring tube exciter coils (inlet or outlet) is likely defective
3-4
386
3-5
387
3-6
388
Amplifier error
3-7
389
4-0
390
4-1
502
Upload and download of device files. Currently no other commands are
possible.
4-2
586
The fluid properties do not allow normal measuring operation.
4-3
587
Extreme process conditions exist.
The measuring system can therefore not be started.
4-4
588
Overdriving of the internal analog to digital converter.
A continuation of the measurement is no longer possible!
4-5
601
Positive zero return active
Содержание HART Proline Promass 83
Страница 2: ...Proline Promass 83 2 Endress Hauser...
Страница 147: ...Proline Promass 83 Endress Hauser 147...
Страница 148: ...www addresses endress com...