Commissioning
Proline Promass 83
66
Hauser
6.3.4
"Gas Measurement" Quick Setup menu
The measuring device is not only suitable for measuring liquid flow. Direct mass
measurement based on the Coriolis principle is also possible for measuring the flow rate of
gases.
!
Note!
• Before carrying out the Quick Setup "Gas measurement" the Quick Setup "Commissioning"
has to be executed
• Only mass and Corrected volume flow can be measured and output with the gas
measurement mode. Note that direct density and/or volume measurement is not possible!
• The flow ranges and measuring accuracy that apply to gas measurement are not the same
as those for liquids.
• If corrected volume flow (e.g. in Nm3/h) is to be measured and output instead of the mass
flow (e.g. in kg/h), change the setting for the CORRECTED VOLUME CALCULATION
function to "FIXED REFERENCE DENSITY" in the "Commissioning" Quick Setup menu.
Corrected volume flow can be assigned as follows:
– to a display line,
– to the current output,
– to the pulse/frequency output.
Performing the "Gas Measurement" Quick Setup
This Quick Setup menu guides you systematically through the setup procedure for all the
device functions that have to be parameterized and configured for gas measurement.
a0002618-en
Fig. 38:
"Gas Measurement" Quick Setup menu
Recommended settings are found on the following page.
+
+
E
1004
B
6400
6402
6403
E
Esc
E
+
-
XXX.XXX.XX
Setup
Gas measurement
Quick
Setup
Assign
Low flow cut off
On value
Low flow cut off
Off value
Low flow cut off
HOME
POSITION