Proline Promass 83
Technical data
Hauser
131
10.7
Installation
Installation instructions
Inlet and outlet runs
There are no installation requirements regarding inlet and outlet runs.
Length of connecting cable
remote version
max. 20 m (65 ft)
System pressure
10.8
Environment
Ambient temperature range
Sensor and transmitter:
• Standard: –20 to +60 °C (-4 to +140°F)
• Optional: –40 to +60 °C (-40 to +140°F)
!
Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic
regions.
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be
impaired.
Storage temperature
–40 to +80 °C (–40 to +175 °F), pref20 °C (+68 °F)
Degree of protection
Standard: IP 67 (NEMA 4X) for transmitter and sensor
Shock resistance
According to IEC 60068-2-31
Vibration resistance
Acceleration up to 1 g, 10 to 150 Hz, following IEC 60068-2-6
CIP cleaning
Yes
SIP cleaning
Yes
Electromagnetic
compatibility (EMC)
To IEC/EN 61326 and NAMUR Recommendation NE 21