Troubleshooting
Proline Promass 83
90
Hauser
351
to
354
S: CURRENT RANGE n
!: # 351 to 354
S
Current output:
The actual value for the flow lies outside
the set limits.
1. Change the upper or lower limit setting, as
applicable
2. Increase or reduce flow, as applicable
355
to
358
S: FREQ. RANGE n
!: # 355 to 358
S
Frequency output:
The actual value for the flow lies outside
the set limits.
1. Change the upper or lower limit setting, as
applicable
2. Increase or reduce flow, as applicable
359
to
362
S: PULSE RANGE
!: # 359 to 362
S
Pulse output:
Pulse output frequency is out of range.
1. Increase the setting for pulse weighting
2. When selecting the pulse width, choose a value
that can still be processed by a connected counter
(e.g. mechanical counter, PLC etc.).
Determine the pulse width:
– Version 1: Enter the minimum duration that a
pulse must be present at the connected counter
to ensure its registration.
– Version 2: Enter the maximum (pulse)
frequency as the half "reciprocal value" that a
pulse must be present at the connected counter
to ensure its registration.
Example:
The maximum input frequency of the connected
counter is 10 Hz. The pulse width to be entered is:
a0004437
3. Reduce flow
363
S: CUR IN. RANGE
!: # 363
S
Current input:
The actual value for the current lies
outside the set limits.
1. Change set lower-range or upper-range value.
2. Check settings of the external sensor.
379
to
380
S: FREQ. LIM
$
: # 379 to 380
F
The measuring tube oscillation
frequency is outside the permitted
range.
Causes:
– Measuring tube damaged
– Sensor defective or damaged
Contact your E+H service organization.
381
S: FLUIDTEMP.MIN.
$
: # 381
F
The temperature sensor on the
measuring tube is likely defective.
Check the following electrical connections before you
contact your E+H service organization:
– Verify that the sensor signal cable connector is
correctly plugged into the amplifier board.
– Remote version:
Check sensor and transmitter terminal connections
No. 9 and 10.
382
S: FLUIDTEMP.MAX.
$
: # 382
F
383
S: CARR.TEMP.MIN
$
: # 383
F
The temperature sensor on the carrier
tube is likely defective.
Check the following electrical connections before you
contact your E+H service organization:
– Verify that the sensor signal cable connector is
correctly plugged into the amplifier board.
– Remote version:
Check sensor and transmitter terminal connections
No. 11 and 12.
384
S: CARR.TEMP.MAX
$
: # 384
F
385
S: INL.SENS.DEF.
$
: # 385
F
One of the measuring tube exciter coils
(inlet) is likely defective.
Check the following electrical connections before you
contact your E+H service organization:
– Verify that the sensor signal cable connector is
correctly plugged into the amplifier board.
– Remote version:
Check sensor and transmitter terminal connections
No. 4, 5, 6 and 7.
386
S: OUTL.SENS.DEF.
$
: # 386
F
One of the measuring tube exciter coils
(outlet) is likely defective.
387
S: SEN.ASY.EXCEED
$
: # 387
F
Measuring pipe excitation coil is
probably faulty.
No.
Error message / Type
Status signals
(from the
factory,
only HART 7)
Cause
Remedy (spare part
→
97)
1
2.10 Hz
= 50 ms