Installation
Proline Promass 83
18
Hauser
A0010011
Fig. 11:
Promass I: When lines are pitched in a specific direction and at a specific slope: as per hygienic guidelines
(21 mm/m or approximately 2%). Gravity can be used to achieve complete drainability.
1
The arrow indicates the direction of flow (direction of fluid flow through the pipe).
2
The label shows the installation orientation for horizontal drainability.
3
The underside of the process connection is indicated by a scribed line. This line indicates the lowest point of the eccentric
process connection.
Promass I and P with hygienic connections
(mounting clamp with lining between clamp and instrument)
It is not necessary to support the sensor under any circumstances for operational
performance. If the requirement exists to support the sensor the following recommendation
should be followed.
A0007397
Fig. 12:
Promass P, mounted with mounting clamp
Esc
Esc
E
-
+
1
2
3
21 mm/m ( 2%)
~
0.83 in/3.28 ft ( 2%)
~
A
B
C
DN
8
15
25
40
50
A
298
402
542
750
1019
B
33
33
33
36.5
44.1
C
28
28
38
56
75