Proline Promass 83
Troubleshooting
Hauser
91
388
to
390
S: AMP. FAULT
$
: # 388 to 390
F
Amplifier error
Contact your Hauser service organization.
No. # 5xx
→
Application error
501
S: SW.-UPDATE ACT.
!: # 501
C
New amplifier or communication (I/O
module) software version is loaded.
Currently no other functions are
possible.
Wait until process is finished. The device will restart
automatically.
502
S: UP-/DOWNLOAD ACT.
!: # 502
C
Up- or downloading the device data via
configuration program. Currently no
other functions are possible.
Wait until process is finished.
571
S: BATCH RUNNING
!: # 571
–
Batching is started and active (valves
are open).
No measures needed (during the batching process
some other functions may not be activated).
572
S: BATCH HOLD
!: # 572
–
Batching has been interrupted (valves
are closed).
1. Continue batching with command "GO ON".
2. Interrupt batching with "STOP" command.
586
S: OSC. AMP. LIMIT
$
: # 586
F
1)
The fluid properties do not allow a
continuation of the measurement.
Causes:
– Extremely high viscosity
– Process fluid is very inhomogeneous
(gas or solid content)
Change or improve process conditions.
587
S: TUBE OSC. NOT
$
: # 587
F
1)
Extreme process conditions exist. The
measuring system can therefore not be
started.
Change or improve process conditions.
588
S: GAIN RED.IMPOS
$
: # 588
F
1)
Overdriving of the internal analog to
digital converter.
Causes:
– Cavitation
– Extreme pressure pulses
– High gas flow velocity
A continuation of the measurement is
no longer possible!
Change or improve process conditions, e.g. by
reducing the flow velocity.
No. # 6xx
→
Simulation mode active
601
S: POSITIVE ZERO
RETURN
!: # 601
C
1)
Positive zero return active.
"
Caution!
This message has the highest display
priority.
Switch off positive zero return.
611
to
614
S: SIM. CURR. OUT. n
!: # 611 to 614
C
Simulation current output active.
621
to
624
S: SIM. FREQ. OUT n
!: # 621 to 624
C
Simulation frequency output active.
Switch off simulation.
631
to
634
S: SIM. PULSE n
!: # 631 to 634
C
Simulation pulse output active.
Switch off simulation.
641
to
644
S: SIM. STAT. OUT. n
!: # 641 to 644
C
Simulation status output active.
Switch off simulation.
651
to
654
S: SIM. RELAY n
!: # 651 to 654
C
Simulation relay output active.
Switch off simulation.
No.
Error message / Type
Status signals
(from the
factory,
only HART 7)
Cause
Remedy (spare part
→
97)