Operation
Proline Promass 83
50
Hauser
8-3
359
Pulse output:
Pulse output frequency is out of range.
8-4
360
8-5
361
8-6
362
9-0
379
The measuring tube oscillation frequency is outside the permitted range.
9-1
380
9-2
381
The temperature sensor on the measuring tube is likely defective.
9-3
382
9-4
383
The temperature sensor on the carrier tube is likely defective.
9-5
384
9-6
385
One of the measuring tube exciter coils (inlet or outlet) is likely defective.
9-7
386
10-0
387
10-1
388
Amplifier error
10-2
389
10-3
390
11-6
471
Max. permitted batching time has been exceeded.
11-7
472
Underbatching: the minimum quantity was not reached.
Overbatching: the maximum permitted batching quantity was exceeded.
12-0
473
The predefined batch quantity point was exceeded.
End of filling process approaching.
12-1
474
Maximum flow value entered is overshot.
12-7
501
New amplifier software version is loaded. Currently no other commands are
possible.
13-0
502
Upload and download of device files. Currently no other commands are
possible.
13-2
571
Batching process in progress (valves are open)
13-3
572
Batching process has been stopped (valves are closed)
13-5
586
The fluid properties do not allow normal measuring operation.
13-6
587
Extreme process conditions exist.
The measuring system can therefore not be started.
13-7
588
Overdriving of the internal analog to digital converter.
A continuation of the measurement is no longer possible!
14-3
601
Positive zero return active
14-7
611
Simulation current output active
15-0
612
15-1
613
15-2
614
15-3
621
Simulation frequency output active
15-4
622
15-5
623
15-6
624
Byte-bit
Error No.
Short error description
→
87.