Commissioning
Proline Promass 83
62
Hauser
6.3.3
"Batching" Quick Setup menu
This Quick Setup menu guides you systematically through the setup procedure for all the
device functions that have to be parameterized and configured for batching operation. These
basic settings allow simple (one step) batching processes.
Additional settings, e.g. for the calculation of after runs or for multi-stage batching
procedures, must be made via the function matrix itself (see the "Description of Device
Functions" manual).
"
Caution!
The "Batching" Quick Setup sets certain device parameters for discontinuous measurement
operation.
If the measuring instrument is used for continuous flow measurement at a later time, we
recommend at you rerun the "Commissioning" and/or "Pulsating Flow" Quick Setup.
!
Note!
• Before carrying out the Quick Setup "Batching" the Quick Setup "Commissioning" has to be
executed.
• This function is only available when the additional "batching" software is installed in the
measuring device (order option). You can order this software from E+H as an accessory at
a later date.
• You can find detailed information on the batching functions in the separate "Description of
Device Functions" manual".
• You can also directly control filling process using the local display. During Quick Setup, an
appropriate dialog appears concerning the automatic display configuration. Acknowledge
this by clicking "YES".
This assigns special batching functions (START, PRESET, MATRIX) to the bottom line of
the display. These can be directly executed onsite using the three operating keys (
O
/
S
/
F
). Therefore, the measuring device can be fully deployed in the field as a "batch
controller".
• The Quick Setup "Batching" is not available for Promass X.