Technical data
Proline Promass 83
112
Hauser
Flow values (example)
Repeatability
• Mass flow and volume flow (liquids): ±0.05% o.r.
• Mass flow (gases): ±0.25% o.r.
• Density (liquids): ±0.00025 g/cm
3
• Temperature: ±0.25 °C ± 0.0025 · T °C; ±0.5 °F ± 0.0015 · (T - 32) °F
Influence of medium temperature
When there is a difference between the temperature for zero point adjustment and the
process temperature, the typical measured error of the sensor is ±0.0002% of the full scale
value / °C (±0.0001% of the full scale value/°F).
Influence of medium pressure
A difference in pressure between the calibration pressure and the process pressure does not
have any effect on the accuracy.
Design fundamentals
Dependent on the flow:
• Flow
≥
Zero point stability ÷ (Base accuracy ÷ 100)
– Max. measured error: ±Base accuracy in % o.r.
– Repeatability: ± ½ · Base accuracy in % o.r.
• Flow < Zero point stability ÷ (Base accuracy ÷ 100)
– Max. measured error: ± (Zero point stability ÷ measured value) · 100% o.r.
– Repeatability: ± ½ · (Zero point stability ÷ measured value) · 100% o.r.
Turn down
Flow
Max. measured error
[kg/h]
[lb/min.]
[% o.r.]
250:1
0.4
0.0147
1.250
100:1
1.0
0.0368
0.500
25:1
4.0
0.1470
0.125
10:1
10
0.3675
0.100
2:1
50
1.8375
0.100
Design fundamentals
Base accuracy for
Mass flow liquids
0.10
Volume flow liquids
0.10
Mass flow gases
0.50