NM2B
REPLACEMENT PROCEDURES (continued)
*
4.10.19.9
The yellow O
2
/N
2
O FLOW RATIO lamp on the alarm
panel should be lighted.
4.10.19.10 Close the oxygen flow control valve. The nitrous oxide
flow should decrease proportionally, and the oxygen
concentration should remain between 21% and 29%
oxygen.
4.10.19.11 The nitrous oxide flow should stop when the flow of
oxygen is between 200 and 300 ml per minute.
4.10.19.12 Close the N
2
O flow control valve and set the System
Power switch to
STANDBY
.
NOTE:
If the ORMC is not working properly,
perform the adjustment procedure given in
Section 5.
4.10.20 Replace the bottom plate of the vapor box. Be sure it is oriented
correctly with the ORMC mounting holes in the correct position.
4.10.21 Secure the ORMC to the bottom of the vapor box with its two
mounting screws.
4.10.22 Replace the rear cover of the vapor box.
4.10.23 Perform the PMS Procedure given in Section 6.
Rev. A
4-33