NM2B
TROUBLESHOOTING GUIDE (continued)
GUIDE 2: Loss of Breathing Pressure Monitor
N
Y
N
Y
N
Y
N
Y
N
Y
START
UNIT IS
FUNCTIONAL
DOES
UNIT PASS
PMS?
PERFORM A
COMPLETE PMS
ON UNIT
CONTACT NAD
SERVICE DEPT.
REINSTALL ORIGINAL
PROCESSOR
ASSEMBLY
REPLACE
PROCESSOR ASSEMBLY
AS OUTLINED IN
PROCEDURE 4.26
CONNECT TUBING TO
TRANSDUCER ON
PROCESSOR BOARD
(ALSO CHECK
CONTINUITY OF
INTERNAL TUBING
CONNECTIONS)
CONNECT
PILOT LINE
AT BOTH ENDS
IS
BREATHING
PRESSURE MONITOR
FUNCTIONAL?
IS
BREATHING
PRESSURE TUBING
CONNECTED TO
TRANSDUCER ON
PROCESSOR
BOARD?
IS
PILOT LINE
CONNECTED TO
INTERFACE PANEL AND TO
PATIENT Y-PIECE OR
ABSORBER?
IS
BREATHING
PRESSURE MONITOR
FUNCTIONAL?
Rev. A
3-5