PMC PROCEDURE (continued)
NM2B
6.14.2
Air O.F.P.D. - if applicable
*
6.14.2.1
Set the gas selector switch to ALL GASES.
6.14.2.2
Open the air cylinder valve - if applicable.
6.14.2.3
Connect the air pipeline hose - if applicable.
6.14.2.4
Open one (1) oxygen cylinder valve.
6.14.2.5
Set the air flow to 4 l/min; set the oxygen flow to 4 l/min.
6.14.2.6
Close the oxygen cylinder valve.
( )
6.14.2.7
Does the flow of air cease when the oxygen pressure is depleted?
___ (Y)
6.14.2.8
Close the air flow control valve.
6.14.2.9
Disconnect the air pipeline hose - if applicable.
6.14.3
Oxygen-Helium O.F.P.D. - if applicable
*
6.14.3.1
Set the gas selector switch to ALL GASES.
6.14.3.2
Open the oxygen-helium cylinder valve.
6.14.3.3
Open one (1) oxygen cylinder valve.
6.14.3.4
Set the oxygen-helium flow to 4 l/min; set the oxygen flow to 4
l/min.
6.14.3.5
Close the oxygen cylinder valve.
( )
6.14.3.6
Does the flow of oxygen-helium cease when the oxygen pressure is
depleted? ___ (Y)
6.14.3.7
Close the oxygen-helium flow control valve.
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
Rev. Y
6-32
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...