NM2B
ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
SENSOR
CORD
CONNECTOR
RECEPTACLE
ON INTERFACE
PANEL
SV20316
SENSOR HOUSING
SENSOR CAPSULE
SENSOR HOUSING
COVER
INSPIRATORY
VALVE DOME
INSPIRATORY
VALVE DOME
PLUG
Figure 5-6: OXYGEN SENSOR CONNECTIONS AND CALIBRATION SCREEN
CALIBRATION MENU
O2MED
OFFSET O2 CELL A
0138
0127
OFFSET O2 CELL B
0138
0129
BAROMED
OXYGEN OFFSET READINGS STORED
PRESSURE SPAN (60 cm H2O)
0243
0481
EXIT
TRACE
TREND
Rev. E
5-19