TROUBLESHOOTING GUIDE (continued)
NM2B
GUIDE 7: No Oxygen Ratio Monitor Alarms
N
Y
N
Y
N
Y
N
Y
N
Y
N
Y
N
Y
N
Y
START
UNIT IS
FUNCTIONAL
DOES
UNIT PASS
PMS?
PERFORM A
COMPLETE PMS
ON UNIT
CONTACT NAD
SERVICE DEPT.
REINSTALL ORIGINAL
PROCESSOR
ASSEMBLY
ARE O2 / N2O
RATIO ALARMS
FUNCTIONAL?
REPLACE
PROCESSOR ASSEMBLY
AS OUTLINED IN
PROCEDURE 4.26
REINSTALL ORIGINAL
ORMC
IS O2 / N2O
FLOW RATIO
LED INDICATOR
FUNCTIONAL?
REINSTALL ORIGINAL
ALARM CHANNEL
--------------------
REPLACE ORMC
AS OUTLINED IN
PROCEDURE 4.10
IS O2 / N2O
FLOW RATIO
LED INDICATOR
FUNCTIONAL?
REPLACE ALARM
CHANNEL ASSEMBLY
AS OUTLINED IN
PROCEDURE 4.21
IS O2 / N2O
FLOW RATIO
LED INDICATOR
FUNCTIONAL?
CHECK ORMC
ADJUSTMENTS
AS OUTLINED IN
PROCEDURE 5.4
IS
"O2 / N20 LOW"
DISPLAYED ON
CENTRALERT ALARM
PANEL?
IS O2 / N2O
FLOW RATIO
LED INDICATOR
FUNCTIONAL?
ARE O2 / N2O
RATIO ALARMS
FUNCTIONAL?
Rev. A
3-10
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...