ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
NM2B
5.4.31
The nitrous oxide flow should stop when the flow of oxygen is
between 200 and 300 ml per minute.
*
5.4.32
If the O
2
/N
2
O FLOW RATIO lamp turns off when the nitrous oxide
flow drops to between 100 and 200 ml per minute, proceed to Step
5.4.34.
*
5.4.33
If the
O
2
/N
2
O FLOW RATIO
lamp does not turn off when the nitrous
oxide flow drops to between 100 and 200 ml per minute, adjust the
MPL switch until the
O
2
/N
2
O FLOW RATIO
lamp turns off.
5.4.34
Close the N
2
O flow control valve and set the System Power switch to
STANDBY
.
5.4.35
Replace the rear cover of the vapor box.
5.4.36
Perform the PMS Procedure given in Section 6.
Rev. A
5-14