REPLACEMENT PROCEDURES (continued)
WARNING:
When making adjustments to the CRT assembly, use only non-
metallic alignment tools and be careful not to touch any of the
high voltage leads or CRT circuitry.
4.27.11 Turn the System Power switch to ON, and observe the display. If
necessary, adjust the size, centering, contrast, brightness and focus
controls to obtain a sharp and properly aligned display.
NOTE:
Typical locations of CRT adjustments are shown in Figure 4-27A. An
alternate model arrangement is shown in Figure 4-27B.
4.27.12 Turn the System Power switch to STANDBY.
4.27.13 Place the monitor box top shelf into position, connect its ground wire
to the chassis ground tab, and reinstall the top shelf screws.
4.27.14 Perform the PMS Procedure given in Section 6.
Rev. A
4-95